Harmonics (continued)

 

fund. ±3° ... 51st ± 15°

Phase

K-factor (in Amp and Watt)

± 10 %

Sags & Swells

 

 

Recording times (selectable)

4 minutes to 8 days, endless (16 days)

Readings

 

 

V rms actual, A rms actual (cycle by cycle calculation)

.....±(2 % + 10 counts)

V rms max, A rms max

±(2 % + 10 counts)

V rms min, A rms min

±(2 % + 10 counts)

Cursor Readings

 

 

V rms max, A rms max

±(2 % + 12 counts)

V rms average, A rms average

±(2 % + 10 counts)

V rms min, A rms min

±(2 % + 12 counts)

Transients

 

 

Detection of voltage transients

> 40 ns

Useful bandwidth input 1 (with test leads TL24)

DC to 1 MHz

Reference signal

V rms, Hz

After START, the V rms and frequency of the signal are measured.

From these data a pure sinewave is calculated.

 

Detection when transients exceed specified voltage level (selectable)

Voltage levels

20 %, 50 %, 100 %, 200 % of reference signal

Number of transient memories (temporary)

40

Cursor reading

 

 

Vpeak min, Vpeak max at cursor

± 5 % of full scale

Inrush Current

 

 

Graphic display

 

 

Current ranges (selectable)

1 A, 5 A, 10 A, 50 A, 100 A, 500 A, 1000 A

Inrush times (selectable)

1 s, 5 s, 10 s, 50 s, 100 s, 5 min

Cursor readings

 

 

A peak max at cursor 1

± 5 % of full scale

A peak max at cursor 2

± 5 % of full scale

Time between cursors

±(0.2 % + 2 pixels)

38

Page 42
Image 42
Fluke 43B user manual Sags & Swells, Transients, Inrush Current