Fluke 43B Meter, Record, Function Specifications, Diode, Continuity, Capacitance, Temperature

Models: 43B

1 53
Download 53 pages 28.43 Kb
Page 45
Image 45
Meter

 

 

Specifications

 

 

 

Function Specifications

3

Meter

 

 

 

Ohm

 

 

 

Ranges

500.0to 5.000 M, 30.00 M

 

 

±(0.6 % + 5 counts)

Max. Measurement Current

0.5 mA

Measurement Voltage at open circuit

......................................................

< 4V

Diode

 

 

 

Accuracy

±(2 % +5 counts)

Max. Measurement Current

0.5 mA

Measurement Voltage at open circuit

.....................................................

< 4 V

Continuity

 

 

 

Beep

< 30 (± 5 )

Measurement Current

0.5 mA

Detection of shorts

> 1 ms

Capacitance

 

 

 

Ranges

50.00 nF to 500.0 F

 

 

±(2 % +10 counts)

Max. Measurement Current

0.5 mA

Temperature

 

 

 

Ranges (°C or °F)

-100.0 to +400.0 °C or -200.0 to +800.0 °F

 

 

±(0.5 % + 5 counts)

Record

 

 

 

Record times (selectable)

4 min to 8 days, endless (16 days)

Number of readings

1 or 2 simultaneously

Cursor Readings Accuracy

Reading Accuracy ±(2 pixels)

Record is available for the functions:

volts / amps / hertz

power

harmonics

ohms / continuity / capacitance

temperature

scope

41

Page 45
Image 45
Fluke 43B user manual Meter, Record, Function Specifications, Diode, Continuity, Capacitance, Temperature