Fluke 43B user manual Current Probe 80i-500S, Specifications

Models: 43B

1 53
Download 53 pages 28.43 Kb
Page 47
Image 47
Current Probe 80i-500S

Specifications 3

Current Probe 80i-500S

Current Probe 80i-500S

Manual background Safety Characteristics

Designed for measurements on 600 V rms Category III. Pollution degree 2, double or reinforced insulation requirements in accordance with:

EN/IEC 61010-2-032

ANSI/ISA S82

CSA-C22.2 No.1010.1-92

UL1244

Electrical Specifications

 

 

Current range

1A to 500 A rms

AC current over range limit

700 A rms

Maximum 10 minutes, followed by removal from current carrying

 

conductor for 30 minutes.

 

 

Output Signal

1mV AC/A AC

Accuracy

 

 

5 to 10 Hz

 

 

1 to 500 A

-3 dB typically

10 to 20 Hz

 

 

1 to 300 A

±

5 %

300 to 400 A

±

15 %

400 to 500 A

±

25 %

20 to 45 Hz

 

 

1 to 500 A

±

5 %

45 to 65 Hz

 

 

1 to 20 A

± 5 % of reading + 0.3 A

20 to 100 A

± 5 % of reading

 

± 3° phase shift

100 to 500 A

± 2 % of reading

65 Hz to 3 kHz

± 5° phase shift

 

 

1 to 50 A

± (5 % + 0.4 A)

50 to 500 A

±

5 %

Influence of temperature on accuracy

<0.15 % per 10 ° C (18 ° F)

Altitude

 

 

During operation

2.0 km (6560 feet)

While stored

12 km (40 000 feet)

43

Page 47
Image 47
Fluke 43B user manual Current Probe 80i-500S, Specifications