Displays harmonics as a percentage of the fundamental voltage.

Fluke 43B

Users Manual

Table 1. Harmonics- and Power Settings

HARMONICS settings

%r

Displays harmonics as a percentage of the total harmonic voltages (total Vrms value).

%f

Displays harmonics as a percentage of the fundamental voltage.

DC..21 Displays the DC component DC..33 of the signal and 21, 33 or DC..51 51 harmonics.

POWER settings

FUNDAMENTAL

Uses only the fundamental voltage and fundamental current for power calculations.

FULL

Uses the voltage and current of the full frequency spectrum for power calculations.

For signals with harmonics, power readings with FULL selected, will differ from power readings with

FUNDAMENTAL selected.