Fluke 43B

Users Manual

Scope readings

The accuracy of all scope readings is valid from 18 °C to 28 °C with relative humidity up to 90 % for a period of one year after calibration. Add 0.1 x (the specified accuracy) for each °C below 18 °C or above 28 °C. More than one waveform period must be visible on the screen.

V DC, A DC

±(0.5 % + 5 counts)

V AC and V AC+DC (True RMS) input 1

 

DC to 60 Hz

±(1 % + 10 counts)

60 Hz to 20 kHz

±(2.5 % + 15 counts)

20 kHz to 1 MHz

±(5 % + 20 counts)

1

MHz to 5 MHz

±(10 % + 25 counts)

5

MHz to 20 MHz

±(30 % + 25 counts)

A AC and A AC+DC (True RMS) input 2

 

DC to 60 Hz

±(1 % + 10 counts)

60 Hz to 15 kHz

±(30 % + 25 counts)

Frequency (Hz), Pulse width, Duty cycle (2.0 % to 98.0 %)

 

1

Hz to 1 MHz

±(0.5 % + 2 counts)

1

MHz to 10 MHz

±(1 % + 2 counts)

10 MHz to 30 MHz

±(2.5 % + 2 counts)

Phase (Input 1 to Input 2)

 

1

Hz to 60 Hz

±2°

60 Hz to 400 Hz

±5°

Peak voltage

 

Peak max, Peak min

± 5 % of full scale

Peak-peak

± 10 % of full scale

Crest

 

Range

1.0 to 10.0

 

 

±(5 % + 1 counts)