Fluke 43B

Users Manual

Safety Specifications

Safety Characteristics

Designed and tested for measurements on 600 V rms Category III, Pollution Degree 2 in accordance with:

EN/IEC 61010-1 (1993)

ANSI/ISA S82.01-1994

CAN/CSA-C22.2 No.1010.1-92 (including approval)

UL3111-1 (including approval)

Installation Category III refers to distribution level and fixed installation circuits inside a building.

Maximum input voltage Input 1 and 2

 

 

Direct on inputs or with test leads TL24

(see Figure 10)

0 to 66 kHz

600

V rms

> 66 kHz

derating to 5

V rms

With Shielded Banana-to-BNC Adapter Plug BB120

(see Figure 10)

0 to 400 kHz

300

V rms

> 400 kHz

derating to 5

V rms

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 10. Max. Input Voltage v.s. Frequency

 

Maximum floating voltage

 

From any terminal to ground

 

0 to 400 Hz

600 V rms

36

Page 40
Image 40
Fluke 43B user manual Safety Specifications, Maximum input voltage Input 1