Fluke 43B user manual Function Specifications, Electrical functions

Models: 43B

1 53
Download 53 pages 28.43 Kb
Page 41
Image 41
Specifications 3

Specifications 3

Function Specifications

Function Specifications

For all specifications, probe specifications must be added.

Electrical functions

Specifications are valid for signals with a fundamental between 40 and 70 Hz.

Minimum input voltage

4 V peak-peak

Minimum input current

10 A peak-peak (1 mV/A)

Input bandwidth

DC to 15 kHz (unless specified otherwise)

Volts / Amps / Hertz

 

 

 

Readings

V rms (AC+DC), A rms (AC+DC), Hz

Voltage ranges (auto)

5.000 V to 500.0 V, 1250 V

 

 

±(1 % + 10 counts)

Current ranges (auto)

50.00 A to 500.0 kA, 1250 kA

 

 

±(1 % + 10 counts)

Frequency range

10.0 Hz to 15.0 kHz

40.0 to 70.0 Hz

±(0.5 % + 2 counts)

CF Crest Factor range

1.0 to 10.0 ±(5 % + 1 counts)

Power

 

 

 

(1 and 3 Phase, 3 Conductor balanced loads)

 

Readings

Watt, VA, VAR, PF, DPF, Hz

Watt, VA, VAR ranges (auto)

250 W to 250 MW, 625 MW, 1.56 GW

when selected: total (%r):

 

±(2 % + 6 counts)

when selected: fundamental (%f):

±(4 % + 4 counts)

DPF

0.00 to 1.00

0.00 to 0.25

not specified

0.25 to 0.90

 

± 0.04

0.90 to 1.00

 

± 0.03

PF

0.00 to 1.00, ± 0.04

Frequency range

10.0 Hz to 15.0 kHz

40.0 to 70.0 Hz

±(0.5 % + 2 counts)

Harmonics

 

 

 

Number of harmonics

DC..21, DC..33, DC..51

Readings / Cursor readings

 

 

 

V rms / I rms

fund. ±(3 % +

2 counts)

31st ±(5 % + 3 counts), 51st ±(15 % +

5 counts)

Watt

fund. ±(5 % + 10 counts)

31st ±(10 % + 10 counts), 51st ±(30 % +

5 counts)

Frequency of fundamental

 

± 0.25 Hz

37

Page 41
Image 41
Fluke 43B user manual Function Specifications, Electrical functions