Specifications

 

 

2 Specifications

2.1 Specifications

 

 

 

UDA2182 Universal Dual Analyzer

Display

Graphical LCD with white LED Backlight

 

Viewing Area: 66.8 mm (W) X 35.5 mm (H)

 

Dot Pixels: 128 (W) X 64 (H)

 

Display Ranges

pH:

 

 

 

0-14 pH

 

 

 

Temperature: -10 to 110°C (14 to 230°F)

 

ORP:

 

 

 

-1600 to +1600 mV

 

 

Conductivity:

 

 

 

0.01 Cell:

0-2 uS/cm displayable to 200 uS/cm; 0-0.2 mS/cm;

 

 

0-2,000 ppb TDS; 0-200 ppm TDS

 

0.1 Cell:

0-20 uS/cm displayable to 2000 uS/cm; 0-2 mS/cm,

 

 

0-2,000 ppb TDS; 0-2,000 ppm TDS,

 

1.0 Cell:

0-200 uS/cm displayable to 20,000 uS/cm; 0-20 mS/cm;

 

 

0-200 ppm TDS; 0-20 ppt TDS

 

10 Cell:

0-2,000 uS/cm displayable to 99999 uS/cm; 0-200 mS/cm;

 

 

0-2,000 ppm TDS; 0-200 ppt TDS

 

25 Cell:

0-20,000 uS/cm displayable to 99999 uS/cm; 0-500 mS/cm;

 

 

0-10% Concentration displayable to 20%

 

50 Cell:

0-20,000 uS/cm displayable to 99999 uS/cm; 0-1,000 mS/cm;

 

 

0-20% Concentration

 

Temperature: 0 to + 140°C (32 to 284°F)

 

Dissolved Oxygen:

 

 

 

0 - 20 ppm

 

 

 

0 –200 ppb, displayable to 20000 ppb

 

0 – 100% saturation, displayable to 200% saturation

 

Temperature: 2 – 60°C (35.6 – 104°F), must not freeze

Keypad

10 Button Membrane Switch w/Directional Functionality

 

UV/Solvent/Abrasion Resistant

 

Case Material

GE Valox® 357 (un-reinforced thermoplastic polyester)

Performances (Under

Accuracy: 0.5% of reading

 

reference operating

Output Accuracy: +/- 0.01 mA

 

conditions)

Drift: Negligible

 

 

 

Repeatability: 0.05%

 

 

 

Temperature Accuracy:

 

 

pH and Conductivity Thermistor: +/- 0.1°C from –10 to 100° C, +/- 1.0° C from 101° to 140° C

 

pH 1000 ohm RTD: +/- 0.4° C

 

 

D.O. Thermistor: +/- 0.1° C from 0 to 60° C

 

Reference Operating Conditions:

25 +/- 1° C; 10-40% RH; 120 or 240 Vac

Operating Conditions

Ambient Temperature

 

 

 

Operating: 0 to 60°C (32 to 140°F)

 

Storage: -30 to 70°C (-22 to 158°F)

 

RH: 5 to 90% max. Non-condensing up to 40°C (104°F). For higher temperatures the RH specification

 

is derated to maintain constant moisture content

 

Vibration:

 

 

 

5-15 Hz disp

 

8 mm pk to pk

 

15-200 Hz accel

2 G

Standard Analog Output

Two 0-20 mAdc or 4-20 mAdc, 750 ohms max., isolated from inputs, ground, and each other,

 

Independently field-assignable to any parameters and ranges.

 

Proportional to user-set output range(s) of selected parameter(s),

Optional Analog Output

One 0-20 mAdc or 4-20 mAdc, 750 ohms max., isolated from inputs, ground, and each other.

 

Independently field-assignable to any parameters and ranges

January 2009

UDA2182 Universal Dual Analyzer Product Manual

5

Page 17
Image 17
Honeywell manual Specifications, UDA2182 Universal Dual Analyzer