Honeywell UDA2182 manual Out 2 Fault

Models: UDA2182

1 258
Download 258 pages 46.92 Kb
Page 73
Image 73

 

 

 

Configuration

 

 

 

 

 

 

 

Out 1 Fault

Output 1 Fault

Output open conditions. This allows an alarm to be triggered if the

 

 

 

 

respective 4-20 mA output opens.

 

 

Out 2 Fault

Output 2 Fault

 

 

 

 

 

 

 

 

 

 

Out 3 Fault

Output 3 Fault

 

 

 

 

 

 

 

 

Hold

Hold

Engages Hold of Analog Inputs

 

 

 

 

 

 

 

Pharm 1 Warn

Pharmacopoeia 1

The Pharma 1 Display ( Section 5.8) outputs digital Warning signal

 

 

 

Warning

whenever the measured conductivity exceeds the Percent Warning

 

 

 

 

Value selected in the “Pharma Op Panel” on the Pharma Display

 

 

 

 

(Stage 1only)

 

 

 

 

 

 

 

Pharm 1 Fail

Pharmacopoeia 1

The Pharma 1 Display ( Section 5.8) outputs digital Failure signal

 

 

 

Failure

whenever one of the following conditions occur:

 

 

 

 

Stage 1 – Measured Conductivity exceeds 100%

 

 

 

 

Stage 1 – Temperature not within range of 0-100 degrees C

 

 

 

 

Stage 2 – Conductivity is 0.1 µS/cm or greater for 5 minutes

 

 

 

 

Stage 3 – pH not within range of 5 – 7pH

 

 

 

 

Stage 2 and 3 – Temperature not within range of 24 – 26 degrees C.

 

 

 

 

 

 

 

Pharm 2 Warn

Pharmacopoeia 2

The Pharma 2 Display ( Section 5.8) outputs digital Warning signal

 

 

 

Warning

whenever the measured conductivity exceeds the Percent Warning

 

 

 

 

Value selected in the “Pharma Op Panel” on the Pharma Display

 

 

 

 

(Stage 1only)

 

 

 

 

 

 

 

Pharm 2 Fail

Pharmacopoeia 2

The Pharma 2 Display ( Section 5.8) outputs digital Failure signal

 

 

 

Failure

whenever one of the following conditions occur:

 

 

 

 

Stage 1 – Measured Conductivity exceeds 100%

 

 

 

 

Stage 1 – Temperature not within range of 0-100 degrees C

 

 

 

 

Stage 2 – Conductivity is 0.1 µS/cm or greater for 5 minutes

 

 

 

 

Stage 3 – pH not within range of 5 – 7pH

 

 

 

 

Stage 2 and 3 – Temperature not within range of 24 – 26 degrees C.

 

 

 

 

 

 

 

PID 1 Alm 1

PID Control 1 Alarm 1

Control Alarms – See Table 6-15 PID Alarms

 

 

 

 

 

 

 

PID 1 Alm 2

PID Control 1 Alarm 2

 

 

 

 

 

 

 

 

PID 2 Alm 1

PID Control 2 Alarm 1

 

 

 

 

 

 

 

 

PID 2 Alm 2

PID Control 2 Alarm 2

 

 

 

 

 

 

 

 

AC 1 Extract

Auto Cycle 1 Probe

Auto Cycle 1 digital output (Cycle Start Source) configuration selection

 

 

 

Extraction

See Table 6-16 Auto Cycling Configuration.

 

 

 

 

 

 

 

AC 1 Rinse

Auto Cycle 1 Probe

Auto Cycle 1 digital output (Cycle Start Source) configuration selection

 

 

 

Rinse

See Table 6-16 Auto Cycling Configuration.

 

 

 

 

 

 

 

AC1 Cal

Auto Cycle 1

Auto Cycle 1 digital output (Cycle Start Source) configuration selection

 

 

 

Calibration Point 1

See Table 6-16 Auto Cycling Configuration.

 

 

 

 

 

 

 

AC 1 Cal 2

Auto Cycle 1

Auto Cycle 1 digital output (Cycle Start Source) configuration selection

 

 

 

Calibration Point 2

See Table 6-16 Auto Cycling Configuration.

 

 

 

 

 

 

 

AC 1 Fail

Auto Cycle 1 Failure

Auto Cycle 1 Failure is active whenever an Auto Cycle 1 failure occurs

 

 

 

 

Auto Cycle 1digital output (Cycle Start Source) configuration selection

 

 

 

 

See Table 6-16 Auto Cycling Configuration.

 

 

 

 

 

 

 

AC 2 Extract

Auto Cycle 2 Probe

Auto Cycle 2 digital output (Cycle Start Source) configuration selection

 

 

 

Extraction

See Table 6-16 Auto Cycling Configuration.

 

 

 

 

 

 

 

AC 2 Rinse

Auto Cycle 2 Probe

Auto Cycle 2 digital output (Cycle Start Source) configuration selection

 

 

 

Rinse

See Table 6-16 Auto Cycling Configuration.

 

 

 

 

 

 

 

AC 2 Cal

Auto Cycle 2

Auto Cycle 2 digital output (Cycle Start Source) configuration selection

 

 

 

Calibration Point 1

See Table 6-16 Auto Cycling Configuration.

 

 

 

 

 

 

January 2009

UDA2182 Universal Dual Analyzer Product Manual

61

Page 73
Image 73
Honeywell UDA2182 manual Out 2 Fault