Input Calibration

exceeding 0.8 volts, a second process (water reduction) begins to occur and the current again rises. To achieve stable results, the probe should be operated within the flat region so that small changes in the probe characteristics result in negligible changes in probe current.

In some industrial wastewater applications, particularly those in petroleum refineries, active gases dissolved in the wastewater can cause this current-voltage characteristic to shift, moving the flat region to other, usually lower, voltages. Also, in some very rare instances, the chemical treatment of boiler water can cause this current-voltage characteristic to shift, moving the flat region to other, usually lower, voltages.

To summarize, the Probe Bias Test automatically varies the probe voltage while displaying the probe current as shown in the figure. At the completion of the test an opportunity to change the bias voltage is provided. Thus, even where significant gaseous contamination might otherwise interfere with the response of the probe to dissolved oxygen, this advanced feature allows the probe to operate.

(If the results of the probe bias test should ever be significantly different from those shown in the figure, Honeywell Service should be consulted.)

Procedure

Table 8-14 Running a Probe Bias Scan

Step

 

Action

 

Screen

1

 

 

 

 

 

 

 

Press

Calibrate

 

CALIBRATION

 

 

 

Input PV Cal

 

 

 

 

 

 

 

Input Temp Cal

 

 

 

 

 

Output Cal

 

 

 

 

 

Cal History

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

 

 

 

 

 

 

 

Press

Enter

 

PV INPUT CAL

 

 

 

 

 

 

 

 

 

 

In 1 DO Cal

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Use

to select

 

 

 

Input 1 or 2 DO Cal

 

 

 

 

 

 

3

 

 

 

 

 

Enter

IN1 DO CAL

 

Press

Air Cal

 

 

 

 

Sample Cal

 

 

 

 

Reset Cal Factor

 

 

 

Pressure Cal

 

 

 

 

Pressure Offset

 

 

 

 

Reset Prs

Offset

 

 

 

Bias Scan

 

 

 

 

 

 

 

 

Bias Volts

 

 

 

 

 

Reset Bias Volts

 

 

 

 

 

 

 