Input Calibration

Step

Action

Screen

 

 

 

4Put the unit in “Hold” mode Remove the electrode from the process.

Rinse the electrode thoroughly with distilled or de- ionized water

5Press Enter

6

Follow the prompts at the top

“Place probe in Sample”

 

and bottom of the screen.

The display will show the

 

 

 

Oxidation Reduction Potential of

 

 

 

the reference solution as

 

 

 

measured by the electrode

 

 

 

system.

 

 

 

 

“Press Enter when stable”

 

 

 

7

Once the reading is stable

“Change to Sample value”

 

Press

Enter

Use

to change the value to

 

match the actual oxidation-

 

 

 

reduction potential of the

 

 

 

reference solution at its current

 

 

 

temperature.

 

 

 

“Enter to save, Exit to cancel”

 

 

 

 

8

Press

Enter

This will standardize the unit.

 

 

 

9Take the unit out of “Hold” and return to the calibration menu.

10

If the calibration fails, an error Error Messages

 

message will be displayed

Refer to Table 12-2.

 

across the bottom stripe of the

 

 

 

screen.

 

January 2009

UDA2182 Universal Dual Analyzer Product Manual

153

Page 165
Image 165
Honeywell UDA2182 manual This will standardize the unit