Input Calibration

Table 8-5 Oxidation-Reduction Potential of Reference Solutions

at Specified Temperature

 

 

pH Buffer Solution

 

Temperature

 

 

(Honeywell Part Number)

20 °C

25 °C

30 °C

 

 

 

 

 

4.01

@ 25 °C (31103001)

267 mV

263 mV

259 mV

 

 

 

 

 

 

6.86

@ 25

°C (31103002)

100 mV

94 mV

88 mV

 

 

 

 

 

 

7.00

@ 25

°C (not available from Honeywell)

92 mV

86 mV

80 mV

 

 

 

 

 

 

9.00

@ 25

°C **(not available from Honeywell)

–26 mV

–32 mV

–39 mV

 

 

 

 

 

 

9.18

@ 25

°C (31103003)

–36 mV

–43 mV

–49 mV

 

 

 

 

 

 

Procedure

Make sure you have selected “PV Type –ORP” in the Inputs configuration – Table 6-5

Refer to Section 6.4.1 – General Rules for Editing.

Table 8-6 Procedure for Calibrating ORP System Using a Reference Solution

Step

 

Action

 

Screen

1

 

 

 

 

 

 

Press

Calibrate

 

CALIBRATION

 

 

 

Input PV Cal

 

 

 

 

 

 

Input Temp Cal

 

 

 

 

 

Output Cal

 

 

 

 

 

Cal History

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Use

to select

 

 

 

Input PV Cal

 

 

 

 

 

 

2

 

 

 

 

Enter

PV INPUT CAL

 

Press

In 1 pH/ORP Cal

 

 

 

 

In 2 Conduc Cal

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Use

to select

 

 

 

Input 1 or 2 pH/ORP Cal

 

 

 

 

 

 

3

 

 

 

 

 

Enter

IN 1 pH/ORP Cal

 

Press

 

 

 

 

Sample Cal

 

 

 

 

 

ORP Offset

 

0.000

 

 

 

Reset ORP Offset

 

 

 

 

 

 

 

 

 

 

 

 

 

 

152

UDA2182 Universal Dual Analyzer Product Manual

January 2009

Page 164
Image 164
Honeywell UDA2182 manual Procedure