Operating the Analyzer

5.7.12Manual Starting/Stopping the Auto Cycle

Pressing Enter on the Auto Cycle Operational Display brings up an operator panel menu that enables you to manually start or stop an auto cycle sequence or place the cycle in Hold, regardless of whether or not the cycle timer is configured. Start cycle is visible when the Auto Cycle is not active and Stop Cycle is visible during an auto cycle.

The objective of a Cycle Stop is to terminate the Auto Cycle by putting the probe back into the process. The selection of Cycle Stop causes the sequencer to proceed directly to the Probe Insert stage, if Probe Transit is enabled, and then to the Resume Delay stage. Selecting Stop Cycle during Probe Insert advances the sequencer to resume delay. Finally, a Stop Cycle selection during Resume Delay will stop the cycle completely.

If enabled, the cycle timer will still trigger a cycle at the configured time provided the sequencer returns to Cycle Stop beforehand. Otherwise, the cycle will execute at the next available time.

The Hold Cycle selection is available in both auto cycle active and inactive states. When inactive, enabling Hold Cycle will hold the sequencer at the beginning of the sequence until released to continue by disabling the Hold Cycle.

Selecting Hold Cycle during an active cycle will suspend the sequencer at the current step as well as the step timer. Disabling Hold Cycle will resume the sequencer and step timer. During cycle Hold, the status bar will show the message “Auto Cycle in hold”.

Table 5-4 Manually Starting/Stopping the Auto Cycle

Press

Action

Enter

to access the Auto Cycle Operator Panel. You will see:

1

AUTO CYCLE 1

3

 

Cycle Stop

12:48:27

 

 

Start Cycle

No

 

1

0 Hold Cycle

No

2

 

 

Next 06.03.29 13:50:24

 

2

 

 

4

Example – AUTO CYCLE 1 Operator Panel

Note: The item “Start Cycle” is replaced with “Stop

Cycle” when the cycle is active.

To access Start Cycle or Hold Cycle

Enter to access the Start Cycle/Hold Cycle selection.

to change the selection from NO to YES.

 

Start Cycle

 

YES

 

 

 

Hold Cycle

 

No

 

 

 

 

 

 

Enter

to Start or Hold the Auto Cycle.

See Table 5-5 for sequencer steps.

January 2009

UDA2182 Universal Dual Analyzer Product Manual

31