Input Calibration

Step

 

Action

 

 

Screen

4

 

 

 

 

 

 

 

Press

Enter

 

CATION PH

 

 

 

Sample Cal

 

 

 

 

 

 

 

 

 

 

 

 

pH Offset

 

0.00

 

 

 

 

 

Rst pH Offset

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Use

to select

 

 

 

Sample Cal

 

 

 

 

 

5

DO NOT Remove the

 

 

 

 

electrode from the

 

 

 

 

process.

 

 

 

 

 

 

6

Once the reading is stable,

Change to Sample Value”

 

press Enter

Use

keys to change the

 

displayed value to match the

 

 

value on the portable meter.

 

 

“Enter to save, Exit to cancel”

 

 

 

 

7

Follow the prompts at the top

“Cal Complete”

 

 

and bottom of the screen.

To recalibrate, press “Enter”.

 

 

 

 

“Enter = recal, Exit = exit”

 

 

 

 

8

If the calibration fails, an error

Error Messages:

 

 

message will be displayed

See Table 12-2.

 

 

across the bottom stripe of the

 

 

 

 

 

 

screen.

 

 

 

 

Make necessary adjustments

 

 

 

 

and re-calibrate.

 

 

 