Configuration

 

 

 

 

 

Sub-menu

Parameter

Selection or Range

Parameter Definition and Restrictions

selection

 

of Setting

 

 

 

Resume Dly Mins

0 to 30.00

Process resume delay in minutes.

 

 

(default = 0.50)

 

 

 

 

 

 

 

6.15.4pH Auto Cycling Configuration Example

The example in Table 6-17 configures the UDA to perform a rinse function once per day, at 8:00 AM, and once per week perform a 1 point Standardization, using 7 buffer. Then once every 4 weeks, perform a complete 2 point Standardize & Slope, using 7 buffer and 4 buffer. Also assume that the sensor is retracted from the sample line during the rinse and cal (not required).

Table 6-17 Example Auto Cycling Configuration for pH

Sub-menu

Parameter

Selection or Range

Parameter Definition and Restrictions

selection

 

of Setting

 

Auto Cycle 1

Auto Cycling

Enable

Allows auto cycling to be selected. This

Auto Cycle 2

 

 

should be enabled after configuration is

 

 

 

complete.

 

 

 

 

 

Hold Active

Enable

When enabled, the output(s) sourced by input

 

 

 

n for Auto Clean n is in hold during auto

 

 

 

cycling.

 

 

 

 

 

Probe Transit

Enable

When enabled, allows probe extract and

 

 

 

probe insert sequence steps to occur and

 

 

 

automated probe extract and insert

 

 

 

parameters are made available.

 

 

 

 

 

Cycle Start Source

None

 

 

 

 

 

 

Extract Wait Src

Digital Signal 1

This is the end of travel “out” switch on the

 

 

See Table 6-4

extraction device

 

 

 

 

 

Insert Wait Src

Digital Signal 2

This is the end of travel “in” switch on the

 

 

See Table 6-4

extraction device

 

 

 

 

 

Cycle Interval

Daily

Frequency of Auto Cycle to occur daily

 

 

 

 

 

Start Hour

8

Cycle to start at 8:00 AM

 

 

 

 

 

Start Mins

0

 

 

 

 

 

 

Rinse Cycle Cnt

1

Rinse to occur every cycle.

 

 

 

 

 

Cal 1 Cycle Cnt

7

Standardize occurs once every 7 cycles, or once

 

 

 

per week.

 

 

 

 

 

Cal 2 Cycle Cnt

28

Slope Cal occurs once per 28 days, or every 4

 

 

 

weeks

 

 

 

 

 

Max Transit Mins

0.5

If extraction takes longer than 30 seconds, then get

 

(Probe Transit =

 

a “AUTOCYCLE FAIL ALARM”

 

 

 

 

Enabled)

 

 

 

 

 

 

 

Rinse Mins

2

Each rinse duration is 2 minutes

 

 

 

 

January 2009

UDA2182 Universal Dual Analyzer Product Manual

103

Page 115