Honeywell UDA2182 manual Auto Cycling Configuration 16 Auto Cycling Configuration, Offdefault

Models: UDA2182

1 258
Download 258 pages 46.92 Kb
Page 113
Image 113

Configuration

6.15.3 Auto Cycling Configuration

Table 6-16 Auto Cycling Configuration

Sub-menu

Parameter

Selection or Range

Parameter Definition and Restrictions

selection

 

of Setting

 

Auto Cycle 1

Auto Cycling

Disable (default)

Allows auto cycling to be selected. This

Auto Cycle 2

 

Enable

should be enabled after configuration is

 

 

 

complete.

 

Hold Active

Enable (default)

When enabled, the output(s) sourced by input

 

 

Disable

n for Auto Clean n is in hold during auto

 

 

 

cycling.

 

 

 

 

 

Probe Transit

Disable (default)

When enabled, allows probe extract and

 

 

Enable

probe insert sequence steps to occur and

 

 

 

automated probe extract and insert

 

 

 

parameters are made available.

 

 

 

 

 

Cycle Start Source

None or Any Digital

Starts Auto Cycle on specific Digital Signal

 

 

Signal

selected changing from 0 to 1.

 

 

See Table 6-4

 

 

Extract Wait Src

None or Any Digital

Allows selection of a specific Digital Signal

 

(Probe Transit =

Signal

that causes a delay in the probe extraction

 

Enabled)

See Table 6-4

sequence. While the selected digital input is

 

 

 

active the probe extraction will not end unless

 

 

 

a timeout occurs as determined by the

 

 

 

duration configured in Probe Transit Mins. If

 

 

 

a source is configured, the timeout results in

 

 

 

an Auto Cycle Fail. If a digital signal is not

 

 

 

available, the source may be left at “None”

 

 

 

and the extract step will occur for the duration

 

 

 

of Probe Transit Mins.

 

 

 

 

 

Insert Wait Src

None or Any Digital

Allows selection of a specific Digital Signal

 

(Probe Transit =

Signal

that causes a delay in the probe insertion

 

Enabled)

See Table 6-4

sequence. While the selective digital input is

 

 

 

active the probe insertion will not end unless a

 

 

 

timeout occurs as determined by the duration

 

 

 

configured in Probe Transit Mins. If a source

 

 

 

is configured, the timeout results in an Auto

 

 

 

Cycle Fail. If a digital signal is not available,

 

 

 

the source may be left at “None” and the

 

 

 

extract step will occur for the duration of

 

 

 

Probe Transit Mins.

 

 

 

 

 

Cycle Interval

Off(default)

Frequency of Auto Cycle occurrence

 

 

Monthly

 

 

 

Weekly

 

 

 

Daily

 

 

 

Custom

 

 

 

 

 

 

Start Time

Disable (default)

Set specific time for Auto-Cycle to start.

 

(Custom)

Enable

 

 

 

 

 

 

 

 

January 2009

UDA2182 Universal Dual Analyzer Product Manual

101

Page 113