Operating the Analyzer

5.7.13Auto Cycle Fail

The status message “Auto Cycle n Fail” is displayed during a fail state. Once detected, the current cycle proceeds immediately to the Probe Insert step (if enabled) or to the Resume Delay step. The fail state remains for the duration of the Resume Delay, whereupon the fail state returns to zero and the fail message is cancelled.

A fail state also provides a detail message in the lower half of the Auto Cycle display regarding the specific reason for the error. These messages are listed in Section 12.4.

The digital output “AC n Fail” is also available and is active whenever an auto cycle failure has occurred. The auto cycle digital outputs: AC n Extract, AC n Rinse, AC n Cal Pt 1 and AC n Cal and AC n Cal 2 are available as relay digital input sources (See Table 6-4) to control the operation of valves and solenoids for exposure to rinse water and buffer solutions or air to accomplish the automatic probe rinse and calibration functions.

5.7.14 Conditional Sequencer Steps

Table 5-5 Conditional Sequencer Steps for Auto Cycle

Step

AC n

AC n

AC n

AC n

 

Extract

Rinse

Cal

Cal 2

 

 

 

 

 

Cycle Stop

0

0

0

0

Cycle Start

0

0

0

0

(transitional)

 

 

 

 

 

 

 

 

 

Probe Extract

1

0

0

0

Probe Rinse

1

1

0

0

 

 

 

 

 

Cal Point 1

1

1

1

0

 

 

 

 

 

Cal Point 2

1

1

0

1

 

 

 

 

 

Probe Insert

0

1

0

0

 

 

 

 

 

Resume Delay

0

0

0

0

 

 

 

 

 

Output

Hold

State

(if

enabled)

Inactive

Active

Active

Active

Active

Active

Active

Active

Condition

Cycle inactive

Operator panel Cycle Start is Yes or Cycle Start Src state is 1

or cycle timer engages

Probe Transit is enabled

Rinse Cycle Cnt > 0 and enables current cycle

and if Probe Transit enabled, Extract Wait Src is None or state is 1 or Max Transit Timer expires

Rinse timer expires if enabled

and Cal 1 (Cal) Cycle Cnt > 0 and enables current cycle

Rinse timer expires or Cal 1 complete

and Cal 2 Cycle Cnt > 0 and enables current cycle

Auto Cycle Fail

or Probe Transit is enabled

Auto Cycle Fail

or Rinse timer expires or Cal 1 or Cal 2 complete

and Insert Wait Src is None or state is 1 or Max Transit Timer expires

32

UDA2182 Universal Dual Analyzer Product Manual

January 2009

Page 44
Image 44
Honeywell UDA2182 manual Auto Cycle Fail, Step AC n Extract Rinse Cal, Output Hold State Enabled, Condition