Honeywell UDA2182 manual Default = 77ºF, Default = 25ºC, None default

Models: UDA2182

1 258
Download 258 pages 46.92 Kb
Page 76
Image 76

Configuration

Sub-menu

Parameter

Selection or

 

Parameter Definition

selection

 

Range of

 

 

 

 

 

 

Setting

 

 

 

 

 

Temp Deg F or C

14.0 to 230.0ºF

Temp Deg F or C will appear depending on what

 

(Temp Type =

default = 77ºF

Temperature Unit was selected in “Maintenance”

 

Manual)

-10 to 110ºC

setup group, parameter “Temp Units”.

 

 

 

 

 

 

 

 

default = 25ºC

 

 

 

 

 

 

 

 

 

 

 

 

Solu Temp Comp

None (default)

Enter “Solution pH/ºC” value

 

(Not ORP)

Custom

 

H20

Pure Water

 

 

 

 

 

 

 

 

NH3

Ammonia

 

 

 

 

Phosphate

Phosphate

 

 

 

 

Morpholine

Morpholine

 

 

 

 

 

 

 

Solution pH/ºC

0.000 (default)

Measured pH is displayed and transmitted normalized

 

(Solu Temp

to

to a solution temperature of 25°C as determined by

 

-0.050

the current Solution Temperature Coefficient. This is

 

Comp = Custom)

 

 

expressed in units of pH/°C with precision to the

 

 

 

 

(Not ORP)

 

hundredths decimal place. The parameter “Solu Temp

 

 

 

Coeff” allows the selection of the following entries.

 

 

 

Follow the “General Rules for Editing” in section

 

 

 

6.4.1 to make the changes. (-) Will appear when first

 

 

 

digit to the right of decimal point is changed.

 

 

 

 

Solution Type

Temp Coefficient

 

 

 

 

 

 

 

 

 

 

 

None (Default)

0.000

 

 

 

 

 

H2O (Pure Water)

-0.016

 

 

 

 

 

NH3 (Ammonia)

-0.032

 

 

 

 

 

PO4 (Phosphate)

-0.032

 

 

 

 

 

C4H9NO (Morpholine)

-0.032

 

 

 

 

 

Custom

User Entry

 

 

PV Bias

-99999 to

PV Bias Constant - is used to compensate the input

 

 

99999

for drift of an input value.

 

 

 

 

default = 0.00

 

 

 

 

 

 

 

 

 

Failsafe

-99999 to

The output value to which the output will go to protect

 

 

99999

against the effects of failure of the equipment.

 

 

default = 14.00

 

 

 

 

 

 

 

 

 

Filter Time

0 to 120

A software digital filter is provided for dampening the

 

 

default = 0

process noise. This filter is applied before the limit

 

 

 

functions.

 

 

 

 

 

 

 

 

 

64

UDA2182 Universal Dual Analyzer Product Manual

January 2009

Page 76
Image 76