Input Calibration

Step

 

Action

 

Screen

4

 

 

 

 

 

Press

Enter

 

IN 1 pH/ORP Cal

 

 

 

 

 

Auto Buffer Cal

 

 

 

 

Buffer Cal

Sample Cal

Buffer Group

pH Offset

pH Slope

Reset pH Offset

Reset pH Slope

 

 

Use

to select

 

 

 

Sample Cal

 

 

 

 

5

Put the unit in “Hold”

 

 

 

mode

 

 

 

DO NOT Remove the

 

 

 

electrode from the

 

 

 

process.

 

 

 

 

 

6

Press Enter

“Place probe in Sample”

 

The display will show the pH of

 

Follow the prompts at the top

the process as measured by the

 

and bottom of the screen.

electrode system.

 

 

“Press Enter when stable”

7Collect a beaker of the process sample from a point near the electrode mounting and measure its pH value with a calibrated portable instrument.

Special instructions for high-purity water applications

For a high purity water application, do not remove the sample from the process for measurement. Bring the portable instrument to the sampling site and measure a continuously flowing sample that has not been exposed to air. This prevents lowering the sample pH by absorption of carbon dioxide from the air.

8

Once the reading is stable,

Change to Sample Value”

 

press Enter

Use

keys to change the

 

displayed value to match the

 

 

value on the portable meter.

 

 

“Enter to save, Exit to cancel”

 

 

 

9

If the calibration fails, an error

Error Messages:

 

message will be displayed

See Table 12-2.

 

across the bottom stripe of the

 

 

 

 

screen.

 

 

 

Make necessary adjustments

 

 

 

and re-calibrate

 

 

 

 

 

 

January 2009

UDA2182 Universal Dual Analyzer Product Manual

149