Input Calibration

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Step

 

Action

 

 

 

 

 

Screen

 

 

 

 

 

 

 

 

 

 

Use

to select

 

 

 

 

 

 

 

 

 

 

Bias Scan

 

 

 

 

 

 

 

 

 

 

 

 

 

4

Press

Enter

to initiate the

 

 

 

 

You will see:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Bias Scan screen

 

 

 

 

 

IN1 BIAS SCAN

 

 

 

 

 

 

 

Enter to scan

0.55V

144μA

 

 

 

 

 

 

 

240

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

160

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

80

 

 

 

 

 

 

 

 

 

 

 

At any time press “Exit” to

 

 

 

 

 

 

 

 

 

 

 

 

 

 

μA 00

0.2 0.4

0.6

0.8 1V

 

 

 

abort scan.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

µA may be 0, 40, 80, 120

 

 

 

 

 

 

 

 

 

 

 

5

Press

Enter

to start scan

Scan in Progress (Example)

 

 

The bias voltage is adjusted down

 

 

 

 

 

 

 

 

 

 

from its original value (usually 0.55V)

 

 

 

 

 

at 25mV/sec until 0V is reached.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

IN1 BIAS SCAN

 

 

 

 

 

 

 

 

 

 

Scanning

0.05V

13μA

 

 

 

 

 

 

 

 

240

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

160

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

80

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

μA 0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

0.2 0.4

0.6

0.8 1V

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Then the bias voltage is driven up to 1.0 V at 25 mV/sec until “Scan complete” appears.

IN1 BIAS SCAN

Scan complete

0.90V

236μ

A

240

 

 

 

 

 

160

 

 

 

 

 

80

 

 

 

 

 

μA 0

 

 

 

 

 

0 0.2 0.4

0.6

0.8 1V

 

 

 

 

 

 

and finally, it is driven down again until it has returned to the value it had just before the test was initiated. During this voltage sweep, the probe current is monitored and the graph of current as a function of voltage is displayed.

IN1 DO BIAS SCAN

Ent=save,=Chng 0.55V 144μA 240

160

80

μA 00 0.2 0.4 0.6 0.8 1V

At the completion of the test an opportunity to change the bias voltage is provided.

Press to change the bias voltage, or

January 2009

UDA2182 Universal Dual Analyzer Product Manual

175

Page 187