Calibration History

11 Calibration History

11.1Overview

Calibration History records every successful input or output calibration with timestamp, with detail available on cal type and before and after cal values by scrolling and selecting cal event name. Calibration records are listed top down from most recent to least recent. Each line in the list consists of a calibration event name and the date and time of occurrence.

Successful automatic cals from auto cycling also recorded and identified. Status warns of cal history at 50% and 90% and when erasing old records.

Accessing the Main Calibration Menu and sub-menus

Press Calibrate . The Main Calibration Menu will appear.

CALIBRATION

Input PV Cal

Input Temp Cal

Output Cal

Cal History

Use the keys to highlight the “Cal History” selection. Press “Enter”.

Calibration Records

Table 11-1 Cal History items

Item

 

Values

Calibration event

In 1 PV Cal

name

In 2 PV Cal

 

In 1 Temp Cal

 

In 2 Temp Cal

 

Out 1 4mA Cal

 

Out 1 20mA Cal

 

Out 2 4mA Cal

 

Out 2

20mA Cal

 

Out 3

4mA Cal

 

Out 3

20mA Cal

 

 

 

Item

Values

Calibration type

Sample Conduc

 

Sample pH/ORP

 

Buffer pH 1

 

Buffer pH 2

 

Auto Buffer pH 1

 

Auto Buffer pH 2

 

Auto Cycle pH 1

 

Auto Cycle pH 2

 

Sample DO

 

Auto Air DO

 

Auto Cycle DO

The list is fully scrollable and individual records are selectable for further detail by highlighting specific event names and pressing Enter. The calibration history has a capacity of 128 records.

January 2009

UDA2182 Universal Dual Analyzer Product Manual

189