Input Calibration

8.5.5Viewing and resetting pH Offset and (Standardization) pH Slope

If the calibration is suspect, you can reset the pH Offset and pH Slope and calibrate again.

In the same screen as “Sample Cal”, use the keys to highlight “Reset pH Offset” or “Reset pH Slope”.

IN 1 pH/ORP Cal

Auto Buffer Cal

Buffer Cal

Sample Cal

Buffer Group

pH Offset

(Read only)

pH Slope

(Read only)

Reset pH Offset

 

Reset pH Slope

 

 

 

Figure 8-1 Resetting pH Offset and pH Slope

Press ENTER. The pH Offset or pH Slope will be reset to (Factory calibration default for Durafet III).