Input Calibration

Step

Action

Screen

 

 

 

5Press Enter

6

Follow the prompts at the top

“Place probe in Sample”

 

and bottom of the screen.

The display will show the

 

 

 

conductivity of the reference

 

 

 

solution as measured by the cell

 

 

 

and Analyzer system.

 

 

 

“Press Enter when stable”

 

 

 

7

Once the reading is stable,

“Change to Sample value”

 

Press

Enter

Use

to change the value to

 

match the actual conductivity of

 

 

 

the reference solution at its

 

 

 

current temperature.

 

 

 

“Enter to save, Exit to cancel”

 

 

 

 

8

Press

Enter

This will save the Calibration

 

Trim Value. If the calibration trim

 

 

 

adjustment is successful, the

 

 

 

calibration menu will again be

 

 

 

displayed.

 

 

 

Return the cell to the process.

 

 

 

Repeat the operation for the

 

 

 

other cell.

 

 

 

9

If the calibration fails, an error

Error Messages

 

message will be displayed

Refer to Table 12-2.

 

across the bottom stripe of the

 

 

 

 

screen.

 

 

 