Honeywell UDA2182 manual Dissolved Oxygen, Ground screw

Models: UDA2182

1 258
Download 258 pages 46.92 Kb
Page 143
Image 143

Inputs and Outputs Wiring

7.7 Dissolved Oxygen

 

Wire

 

Signal

Cable shield (Blue)

Color

 

Name

 

 

 

to chassis ground screw

Clear

10

Cathode

 

Green

9

Reference

 

Red

8

Anode

 

Black*

7

Guard

 

 

6

 

 

Yellow

5

RTH Low

 

Orange

4

RTH High

 

 

3

 

 

 

2

 

Wire to chassis

1

Earth Ground

ground screw

 

 

*Older Dissolved Oxygen probes may have a White/Black Guard wire instead of a Black Guard wire.

Some cables have connectors on the leads. Cut off the connectors, skin and tin the leads and then wire to the screw terminals on the boards

Figure 7-16 Terminal Designations for Dissolved Oxygen with Integral Cable

CAUTION

When installing the probe, the wiring must be done in the order shown below even if the analyzer is not powered. This is because the DO Input card is continuously supplying a voltage (bias) to the terminals.

Connecting – Blue Shield wire first, then in this order:

Red

Green

Coax (clear)

Guard (Black)

Yellow

Orange

Disconnecting – Go in reverse

Orange – first

Yellow

Guard (Black)

Coax (clear)

Green

Red

Blue Shield Wire

January 2009

UDA2182 Universal Dual Analyzer Product Manual

131