Honeywell UDA2182 manual Output Calibration, Required equipment

Models: UDA2182

1 258
Download 258 pages 46.92 Kb
Page 191
Image 191

Outputs Calibration

9.2 Output Calibration

Introduction

The UDA2182 is available with two standard and one optional analog outputs. The output signals can be adjusted to trim the high and low output current or voltage values over a range of ± 0.4 % of span to compensate for component tolerance variations.

Accessing the Main Calibration Menu and sub-menus

Press Calibrate . The Main Calibration Menu will appear.

CALIBRATION

Input PV Cal

Input Temp Cal

Output Cal

Cal History

Use the keys to highlight the “Output Cal” selection.

Press Enter to display the sub-menu for that selection.

Required equipment

Output calibration involves connecting a meter to the Analyzer’s output terminals. The meter required for output calibration depends on the type of outputs.

Current outputs: current meter capable of resolving 0.01 mA over the range 0 to 20 mA dc

Voltage outputs: a 250 ohm ± 0.05 % shunt and a volt meter (capable of measuring 1 to 5 Vdc within 1 mV)

A screwdriver to fit the terminal block screws and the screw securing the terminal board retainer is also required.

Procedure

To calibrate outputs, follow the procedure described in Table 9-1 Procedure for Calibrating Analyzer Outputs. The output terminals are inside the case as shown in Figures 6-1 through 6-6.