Input Calibration

Resetting Pressure Offset or Bias Volts

If the calibration is suspect, you can reset any of these values and calibrate again.

In the same screen as “IN 1 DO Cal”, use the keys to highlight “Reset Prs Offset” or “Reset Bias Volts”.

IN1 DO CAL

Air Cal

Sample Cal

Reset Cal Factor

Pressure Cal

Pressure Offset

Reset Prs Offset

Bias Scan

Bias Volts

Reset Bias Volts

Figure 8-6 Resetting Pressure Offset or Bias Volts

Press ENTER. The selected value will be reset to (default).