Input Calibration

8.7.6Resetting Calibration Trim

If the calibration is suspect, you can reset the Calibration Trim and calibrate again.

In the same screen as “Sample Cal”, use the keys to highlight “Reset Trim”.

IN2 Conduc Cal

Sample Cal

Cal Trim1.00

Reset Cal Trim

Figure 8-3 Resetting Calibration Trim

Press ENTER. The Calibration Trim will be reset to 1.00 (default).

162

UDA2182 Universal Dual Analyzer Product Manual

January 2009

Page 174