Input Calibration

Table 8-3 Procedure for Buffering Method of Calibrating pH Electrodes

Step

 

Action

 

Screen

1

 

 

 

 

 

 

Press

Calibrate

 

CALIBRATION

 

 

 

Input PV Cal

 

 

 

 

 

 

Input Temp Cal

 

 

 

 

 

Output Cal

 

 

 

 

 

Cal History

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Use

to select

 

 

 

Input PV Cal

 

 

 

 

 

 

2

 

 

 

 

Enter

PV INPUT CAL

 

Press

In 1 pH/ORP Cal

 

 

 

 

In 2 Conduc Cal

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Use

to select

 

 

 

Input 1 or 2 pH/ORP Cal

 

 

 

 

 

 

 

3

 

 

 

 

 

 

 

Press Enter

 

IN 1 pH/ORP Cal

 

 

 

Auto Buffer

Cal

 

 

 

 

 

 

 

 

 

 

Buffer Cal

 

 

 

 

 

 

 

Sample Cal

 

 

 

 

 

 

 

 

 

 

 

 

 

Buffer Group

 

 

 

 

 

 

pH Offset

 

 

 

 

 

 

pH Slope

 

 

 

 

 

 

Reset pH Offset

 

 

 

 

 

 

Reset pH Slope

 

 

 

 

 

 

 

 

 

 

 

 

 

Use

to select

 

 

 

 

Buffer Cal

 

 

 

 

 

 

 

 

4

Put the unit in “Hold”

 

 

 

 

 

 

 

mode

 

 

 

 

 

 

Remove the electrode

 

 

 

 

 

 

 

from the process.

 

 

 

 

 

 

Rinse the electrode

 

 

 

 

 

 

 

thoroughly with distilled or

 

 

 

 

 

 

 

de-ionized water

 

 

 

 

 

 

 

 

5

Standardization (adjust

“Place probe in Buffer 1”

 

instrument zero)

The display will show the pH of

 

Press Enter

the buffer 1 solution as

 

measured by the electrode

 

Follow the prompts at the top

system.

 

 

 

“Press Enter when stable”

 

and bottom of the screen.

 

 

 

 

 

 