Honeywell UDA2182 Online Functions Display Details Functions, Process Variable, Values, Bargraphs

Models: UDA2182

1 258
Download 258 pages 46.92 Kb
Page 36
Image 36

Operating the Analyzer

Online Functions

 

Table 5-2 Display Details Functions

 

 

 

 

 

Detail

 

 

Function

Process Variable

When two input boards are installed, the default online screen displays both PVs and

Values

their units of measure, as determined by the input boards, the probe (if memory-

 

embedded) or any measurement configuration options that may be available. When

 

only one input board is installed, the default online screen displays one PV and its units

 

in a larger font size.

 

 

 

The currently selected PV type determines the numerical format and the units of

 

measure on the online PV display. Measured PV is generally displayed in the highest

 

decimal precision possible with five digits and has a potentially displayable range of

 

0.0000 to 99999. The exceptions are dissolved oxygen, pH, ORP and temperature,

 

which are displayed with fixed decimal precision.

 

PV Type determines specific ranges and in the case of Conductivity, cell constant

 

determines available PV Types. Each PV measurement and display is updated every

 

500ms maximum. Each temperature measurement and display is updated every 10

 

seconds maximum.

 

 

 

See the Specific Input configuration for available ranges. (Section 6.6)

 

 

 

 

 

Tag Name

The real-time displays of process values show the instrument’s tag name (or other

 

configurable fixed sixteen-character string) at the top of the screen.

PV Temperature

Each PV value is accompanied by a temperature value for all measurements except

 

ORP, as ORP probes do not contain temperature sensors and no measurement

 

compensation for temperature is required. Temperature values are displayed in units of

 

degrees Fahrenheit or degrees Celsius as determined by configuration.

 

Measured temperature is always expressed in fixed tenths decimal precision and has a

 

displayed range according to input type:

 

 

 

 

 

 

 

PH/ORP

-10.0 to 110.0°C or 14.0 to 230.0°F

 

 

 

 

 

 

 

 

Conductivity

0 to 140.0°C or 32.0 to 284°F

 

 

 

 

 

 

 

 

Dissolved Oxygen

0 to 60.0°C or 32 to 140°F

 

 

 

 

 

 

Status Messages

A text string appears on the bottom of all displays. Online displays provide messages

 

relaying online diagnostics, alarms and other events. Offline screens display messages

 

relevant to data entry and calibration. See Section 12.

 

 

Bargraphs

The Bargraphs will represent up to three output values. The corner indicators represent

 

the physical state of the Relay Outputs [1, 2, 3, and 4].

*Note that all values and indicators on the main (input) display screen are maintained in the input setup group.

24

UDA2182 Universal Dual Analyzer Product Manual

January 2009

Page 36
Image 36
Honeywell UDA2182 manual Online Functions Display Details Functions, Process Variable, Values, Status Messages, Bargraphs