Unpacking

Panel Mounting Dimensions

138

+1

 

 

-0

 

 

[5.43]

+.04

 

 

 

-0

 

 

Panel Cutout

138

+1

-0

 

 

[5.43]

+.04

 

 

 

-0

Customer will need to provide a rear panel support plate to maintain NEMA4 protection if primary panel thickness is less that 2.3mm [0.09”] thick

156

 

33.5

[6.14]

 

[1.32]

 

 

 

CUSTOMER PANEL 1.6[.06] to 6.35 MAX[0.25]

152

[5.98]

156

[6.14]

(4) 22.22[.87] holes for

lead wires and conduit fittings (conduit fittings supplied by user)

Figure 3-1 Panel Mounting Dimensions (not to scale)

Panel Mounting Procedure

Table 3-2 Panel Mounting Procedure

Step

Action

1Mark and cut out the analyzer hole in the panel according to the dimension information in Figure 3-1.

2Orient the case properly and slide it through the panel hole from the front.

Customer will need to provide a rear panel support plate to maintain NEMA4 protection if primary panel thickness is less that 2.3mm [0.09”] thick – See Figure 3-2.

3Remove the mounting kit from the shipping container and clamp the edges of the cutout between the case flange and the supplied U-bracket that is fastened to the rear of the case using (2) M5 X 16mm long screws and (2) M5 lock washers supplied.

January 2009

UDA2182 Universal Dual Analyzer Product Manual

11

Page 23
Image 23
Honeywell UDA2182 manual Panel Mounting Dimensions, Panel Cutout