Configuration

 

 

 

 

 

 

 

Sub-menu

Parameter

 

Selection or

 

Parameter Definition

selection

 

 

Range of

 

 

 

 

 

 

Setting

 

 

 

Input 1 or

For every cell constant the PV type includes selections for both conductivity µS/cm and

Input 2

conductivity mS/cm.

 

 

 

 

Conductivity

Conductivity µS/cm displays µS/cm and provides standard range solution type selections:

 

None, NaCl, Morpholine, HCL, Acid, and NH3.

 

 

 

Conductivity mS/cm displays mS/cm and provides wide range solution type selections: None,

 

HCl, NaCl, H2SO4, and NaOH.

 

 

Upper range limit defaults according to the table below:

For every cell constant the PV type also includes selections for either TDS ppb/TDS ppm or TDS ppm/TDS ppt.:

TDS ppb/ppm provides standard or wide solution type selections and

TDS ppm/ppt provides standard or wide solution type selections. Solution selections are the same as above with the exception of None.

Upper range limit defaults according to the table below:

Cell Const 0.01

Cell Const 0. 1

Cell Const 1

Cell Const 10

Cell Const 25

Cell Const 50

0

- 2 µS/cm

0 - 20 µS/cm

0

- 200 µS/cm

0

- 2000

0 - 20000

0 - 20000

displayable to 200

displayable to

displayable to

µS/cm

µS/cm

µS/cm

µS/cm

2000 µS/cm

20000 µS/cm

displayable to

displayable to

displayable to

0

- 0.2 mS/cm

0 - 2 mS/cm

0

- 20 mS/cm

99999 µS/cm

99999 µS/cm

99999 µS/cm

0

- 200 mS/cm

0 - 500 mS/cm

0 - 1000

0

- 2000 ppb TDS

0 - 20000 ppb

0

- 200 ppm

0

- 2000 ppm

0 - 10 % conc

mS/cm

0

- 200 ppm TDS

TDS

TDS

 

0 - 2000 ppm TDS

0

- 20 ppt TDS

TDS

displayable to

0 - 20 % conc

 

 

 

 

0

- 200 ppt

20%

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TDS

 

 

January 2009

UDA2182 Universal Dual Analyzer Product Manual