Honeywell UDA2182 manual Default = None, Default 10.000, Pharma Timer begins to count down from

Models: UDA2182

1 258
Download 258 pages 46.92 Kb
Page 83
Image 83

 

 

 

Configuration

 

 

 

 

 

Sub-menu

Parameter

Selection or

Parameter Definition

selection

 

Range of

 

 

 

 

Setting

 

 

 

Wire Size AWG +

16 AWG

 

 

 

 

18

 

 

 

 

AWG(default)

 

 

 

 

20 AWG

 

 

 

 

22 AWG

 

 

 

 

 

 

 

 

Wire Size Sq mm +

0.33 to 2.08

 

 

 

 

default = 0.82

 

 

 

 

 

 

 

Pharma Type

None

PhEur - Pharmacopoeia Europa

 

 

PhEur

USP - United States Pharmacopoeia

 

 

USP

 

 

 

 

 

 

default = None

standard procedure stages for determining Purified

 

 

 

Water

 

Pharma PV High

-99999.00 to

Pharma PV High Value – Measured solution

 

 

99999.00

conductivity value scaled for 100%

 

 

(default

 

 

 

 

10.000)

 

 

 

 

 

 

 

Pharma PV Low

-99999.00 to

Pharma PV Low Value - Measured solution

 

 

99999.00

conductivity value scaled for 0%

 

 

(default 0.000)

 

 

 

 

 

 

 

Pharm Tmr Mins

000.0 to 120.0

Pharma Timer Minutes - If the Pharma sample does

 

 

(default

not pass the Stage 1 conductivity requirement a Fail

 

 

10.000)

signal is generated, then the State 2 and Stage 3

 

 

 

tests are conducted. When the Stage 2 or Stage 3 test

 

 

 

is successful, the fail signal is cancelled and the

 

 

 

Pharma Timer begins to count down from the

 

 

 

configured minutes value set here. When the Timer

 

 

 

countdown is completed, the Pharma function block

 

 

 

returns to Stage 1.

 

 

 

 

 

PV Bias

-9999.00 to

PV Bias Constant - is used to compensate the input

 

 

9999.00

for drift of an input value.

 

 

default = 0.000

 

 

 

 

 

 

 

Failsafe

0.0 to 2000

The output value to which the output will go to protect

 

 

default =

against the effects of failure of the equipment.

 

 

2000.000

 

 

 

 

 

 

 

Filter Time

0 to 120.0

A software digital filter is provided for dampening the

 

 

default = 0.000

process noise and is applied before the limit functions.

 

 

 

 

 

January 2009

UDA2182 Universal Dual Analyzer Product Manual

71