Honeywell UDA2182 manual Conductivity of Potassium Chloride Solutions at 25 C

Models: UDA2182

1 258
Download 258 pages 46.92 Kb
Page 172
Image 172

Input Calibration

Table 8-8 Conductivity of Potassium Chloride Solutions at 25 °C

Concentration M*

Conductivity (microSiemens

 

per cm)

0.001

147.0

 

 

0.005

717.8

 

 

0.01

1,413

 

 

0.02

2,767

 

 

0.05

6,668

 

 

* M = Molarity; 1M = 74.555g potassium chloride per liter of solution

Procedure

Table 8-9 Procedure for Performing Calibration Trim Using a Reference Solution

Step

 

Action

 

 

Screen

1

 

 

 

 

 

 

 

 

 

Press

Calibrate

 

 

CALIBRATION

 

 

 

 

Input PV Cal

 

 

 

 

 

 

 

 

Input Temp Cal

 

 

 

 

 

 

Output Cal

 

 

 

 

 

 

Cal History

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

 

 

 

 

 

 

 

 

 

Press

Enter

 

 

PV INPUT CAL

 

 

 

 

 

In 1 pH/ORP Cal

 

 

 

 

 

 

 

In 2 Conduc Cal

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Use to select

Input 1 or 2 Conduc Cal

3

Enter

IN2 Conduc Cal

Press

Sample Cal

 

 

 

 

Cal Trim

 

1.00

 

 

Reset Cal Trim

 

 

 

 

 

 

 

 

 

 

 

4

Remove the cell from the

 

 

process.

 

Rinse the cell thoroughly

 

 

with de-ionized water.

160

UDA2182 Universal Dual Analyzer Product Manual

January 2009

Page 172
Image 172