Input Calibration

8.5.4Sample Method of Calibrating pH Electrodes

Recommended where pH is stable, or for high-purity water applications

This method is recommended only where the pH is stable and changes very slowly. It is also recommended for high-purity water measurement applications. Special instructions for high-purity water applications are provided below.

Materials

To use the sample method, follow the instructions in Table 8-4. Materials required are:

A clean beaker for collecting the sample.

A calibrated portable instrument for measuring pH of the sample.

Distilled or de-ionized water to rinse the electrode.

Procedure

Make sure you have selected “PV Type –pH Glass, pH Durafet, or pH HPW” in the Inputs configuration - Table 6-5.

Refer to Section 6.4.1 – General Rules for Editing.

Table 8-4 Procedure for Sample Method of Calibrating pH Electrodes

Step

Action

Screen

 

 

 

1Prepare the Calibration meter.

2

Calibrate

CALIBRATION

 

 

Press

Input PV Cal

 

 

 

 

 

Input Temp Cal

 

 

 

 

Output Cal

 

 

 

 

Cal History

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Use

to select

 

 

 

Input PV Cal

 

 

 

 

 

 

3

 

 

 

 

 

Enter

PV INPUT CAL

 

Press

In 1 pH/ORP Cal

 

 

 

 

In 2 Conduc Cal

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Use to select

Input 1 or 2 pH/ORP Cal

148

UDA2182 Universal Dual Analyzer Product Manual

January 2009

Page 160
Image 160