Input Calibration

Calibrating a Dissolved Oxygen Probe Using Air Calibration Method

Introduction

This is the simplest and most commonly used method of calibration.

ATTENTION

If “Initial Installation”, power probe and analyzer for 24 hours before first air calibration.

1.Assure that the probe has been powered for at least one hour.

2.Press the Hold button, if required.

3.Expose the probe to air (or air-saturated water) until the temperature and DO value reading stabilizes.

Procedure

Table 8-11 Calibrating a Dissolved Oxygen Probe Using Air Calibration Method

Step

 

Action

 

Screen

1

 

 

 

 

 

 

 

Press

Calibrate

 

CALIBRATION

 

 

 

Input PV Cal

 

 

 

 

 

 

Input Temp Cal

 

 

 

 

 

Output Cal

 

 

 

 

 

Cal History

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

 

 

 

 

 

 

 

Press

Enter

 

PV INPUT CAL

 

 

 

 

 

 

 

 

 

In 1 DO Cal

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Use

to select

 

 

 

Input 1 or 2 DO Cal

 

 

 

 

 

 

3

 

 

 

 

 

Enter

 

IN1 DO CAL

 

Press

Air Cal

Cal

 

 

 

 

Sample

 

 

 

 

Reset Cal Factor

 

 

 

Pressure Cal

 

 

 

 

Pressure Offset

 

 

 

 

Reset Prs Offset

 

 

 

Bias Scan

 

 

 

 

Bias Volts

 

 

 

 

Reset Bias Volts

 

 

 

 

 

 

 

Use to select

Air Cal