Inputs and Outputs Wiring

HPW7000

 

 

Wire

 

 

 

Color

 

 

 

 

15

 

 

Reference cable shield

14

 

 

(White with Green stripe)

13

 

 

to chassis ground screw

 

 

 

 

Reference

 

12

 

 

 

 

cable

 

11

 

 

 

Measurement

 

Clear (center conductor of coax)

10

Measurement

White with Black stripe

 

cable shield

9

 

 

 

(White with Green stripe)

cable

Clear (center conductor of coax)

8

to chassis

 

Red

7

ground screw

 

Thermistor

Jumper

6

 

 

cable

Black

5

 

 

White

4

Thermistor cable shield

 

3

(White with Green stripe)

 

2

to chassis ground screw

 

1

 

 

 

Signal

Name

Reference

Guard

Glass (or ORP)

Counter

RTH 3rd Wire

RTH Low

RTH High

Some cables have connectors on the leads. Cut off the connectors, skin and tin the leads and then wire to the screw terminals on the boards

Figure 7-8 Terminal Designations for HPW7000 System