Honeywell UDA2182 manual Inputs Configuration, Accessing Inputs Menu, Input Configuration

Models: UDA2182

1 258
Download 258 pages 46.92 Kb
Page 75
Image 75

Configuration

6.6 Inputs Configuration

Overview

This group lets you select pH/ORP, Preamp pH, Conductivity, or Dissolved Oxygen Input type and the associated output parameters.

Accessing Inputs Menu

Press Setup to display the Main menu.

Use the keys to select “Inputs” then press Enter to enter the sub-menus.

Input 1 and Input 2 – Direct pH/ORP, Preamp pH, Conductivity, or Dissolved Oxygen are available for selection. Select PV type, read the range, select Temp Type, Solution Temp Compensation, Bias, Failsafe and Filter Time.

For Dissolved Oxygen, also select the Salinity type and Pressure type.

Press to highlight the desired menu selection then press Enter to display the group of parameters.

Refer to “Section 6.4.1 – ”General Rules for Editing”.

Table 6-5 Input Configuration

 

Sub-menu

Parameter

Selection or

Parameter Definition

 

selection

 

Range of

 

 

 

 

Setting

 

 

Input 1 or 2

PV Type

pH Glass

The PV type determines the numerical format and the

 

Direct pH

 

pH HPW

units of measure on the online PV display. Measured

 

ORP

 

pH Durafet

PV is generally displayed in the highest decimal

 

 

(default)

precision possible to .001 and has a potentially

 

 

 

ORP

displayable range of 0.000 to 99999. The exceptions

 

 

 

 

are dissolved oxygen, pH, ORP and temperature,

 

 

 

 

which are displayed with fixed decimal precision. PV

 

 

 

 

Type determines specific ranges.

 

 

 

 

 

 

 

PV Range

0.0 to 14.0 pH

Read Only

 

 

 

-1600 to 1600

 

 

 

 

ORP

 

 

 

 

 

 

 

 

Temp Input

Enable

Enable to allow “Temp Type” selection – see below.

 

 

(ORP only)

Disable

 

 

 

 

 

 

 

 

Temp Type

8550Ω Therm

8550Ω Thermistor

 

 

 

(default)

1000Ω Resistance Temperature Detector

 

 

 

1000Ω RTD

 

 

 

Manual

Manual

 

 

 

 

 

January 2009

UDA2182 Universal Dual Analyzer Product Manual

63

Page 75
Image 75
Honeywell UDA2182 manual Inputs Configuration, Accessing Inputs Menu, Input Configuration