6.15 Auto Cycling Configuration

6.15.1Overview

Auto cycling provides automated timing, control and functionality for the cleaning and calibration of input probes. Each input PV has a dedicated auto cycle function block. The input board type and in the case of pH, the PV type, determines the level of auto cycling capability, as indicated below:

Input Board Type

Auto Cycle Operation

Preamp pH

Rinse, Auto Buffer Cal 1 (zero offset), Auto Buffer Cal 2 (slope)

 

 

pH/ORP, PV Type not

Rinse, Auto Buffer Cal 1, Auto Buffer Cal 2

ORP

 

 

 

pH/ORP, PV type is ORP

Rinse Only

 

 

Conductivity

Rinse Only

 

 

DO ppm

Rinse, Auto Air Cal

 

 

DO ppb

Rinse, Auto Air Cal

 

 

Auto cycling is supported with setup menus, status displays and operational displays (Section 5.7) as well as event (Section 5.11) and calibration history logging (Section 11).

6.15.2Accessing Auto Cycle Menu

Press Setup to display the Main menu.

Use the keys to select “Auto Cycling” then press Enter menu:

Auto Cycle 1 or Auto Cycle 2

Press to highlight the desired menu selection then press group of parameters.

Refer to “Section 6.4.1 – “General Rules for Editing”.

to enter the sub-

Enter to display the

100