Input Calibration

 

 

 

 

 

 

 

Step

Action

 

Screen

 

 

12

Once the reading is stable

“Buffer 2 stability check”

 

 

 

Press Enter

Use

to change the value of

 

 

 

the Buffer.

 

 

 

 

“Up/Down changes Buffer”

 

 

 

 

 

 

 

13

If the calibration fails, an error

Error Messages:

 

 

 

message will be displayed

Buffer span too low

 

 

 

across the bottom stripe of the

 

 

 

 

 

 

 

 

screen.

OFFSET UNDERRANGE

 

 

 

Make necessary adjustments

OFFSET OVERRANGE

 

 

 

and re-calibrate.

Slope underrange

 

 

 

 

 

 

 

 

Slope overrange

 

 

 

 

Solution Unstable

 

 

 

 

Temp Too Low

 

 

 

 

Temp too High

 

 

 

 

See Table 12-2 for definitions

 

 

 

 

 

 

 

8.5.3Buffering Method of Calibrating pH Electrodes

Recommended for most applications

This technique is recommended for best accuracy in most applications.

Materials

Materials required are:

Two standard buffer reference solutions that are at least 2 pH apart from one another.

A container for each, large enough to immerse the electrode to measuring depth.

Distilled or de-ionized water to rinse the electrode.

Procedure

Make sure you have selected “PV Type –pH Glass, pH Durafet, or pH HPW ” in the Inputs configuration - Table 6-5.

Refer to Section 6.4.1 – General Rules for Editing.