Configuration

Sub-menu

Parameter

Selection or Range of

 

 

Parameter Definition

 

selection

 

Setting

 

 

 

 

 

 

Cond Units

NIST (default)

The NIST system of conductivity measurement

 

Type

 

uses units of centimeters, and in the UDA are

 

 

 

specifically µS/cm and mS/cm for conductivity

 

 

 

and KΩ-m for resistivity.

 

 

ISO

The ISO system of conductivity measurement

 

 

 

uses units of meters, and in the UDA are

 

 

 

specifically μS/m, mS/m and S/m for conductivity

 

 

 

and KΩ-m for resistivity.

 

 

 

The conductivity “Units Type” selected here

 

 

 

affects the “PV Type” selections available under

 

 

 

“Setup/Inputs/pH Preamp n. Selection of units

 

 

 

type will scale all live conductivity and Resistivity

 

 

 

readings on monitor and input calibration screens

 

 

 

according to the factors listed:

 

 

 

 

 

 

 

 

 

 

 

 

NIST

ISO

ISO/NIST factor

 

 

 

 

 

μS/cm

μS/m

0.01

 

 

 

 

 

 

 

 

 

 

 

 

 

μS/cm

mS/m

10

 

 

 

 

 

 

 

 

 

 

 

 

 

mS/cm

S/m

10

 

 

 

 

 

 

 

 

 

 

 

 

 

MΩ-cm

KΩ-m

0.1

 

 

 

 

 

 

 

 

 

 

Cond Wire

AWG (default)

“Wire Size Units” allows selection of either AWG

 

Size

Sq mm

or Square millimeters (Sqmm). When changing

 

 

 

units, the wire size parameter value is not

 

 

 

converted. A pop-up message warns you of this.

 

 

 

 

 

Cond Wire Len

Feet (default)

“Wire Length Units” allows selection of either

 

 

Meters

meters or feet for the Wire Length parameter in

 

 

 

the Inputs group. When changing units, the wire

 

 

 

length parameter value is not converted. A pop-

 

 

 

up message warns you of this. If the value is no

 

 

 

longer within range, it will change to closest

 

 

 

range limit.

 

 

 

 

 

 

 

 

 

 

 

Cation Calc

None

None

 

 

 

 

 

pH NH3

Ammonia pH – Specific conductivity

 

 

 

temperature compensation assumes Ammonia

 

 

 

NH3 is the base reagent

 

 

pH Amines

Amine pH – Specific conductivity temperature

 

 

 

compensation assuming a generic amine base.

 

 

ppbCO2

CO2 determination by degassed conductivity.

 

 

 

 

 

 

 

 

110

UDA2182 Universal Dual Analyzer Product Manual

January 2009

Page 122