Temperature Input Calibration

Step

Action

Screen

 

 

 

Use the keys to highlight the desired “Temperature Input” selection.

3

Enter

T1 pH/ORP CAL

Press

 

 

 

Temp Cal

 

 

 

 

Temp Offset

 

(Read only)

 

 

Reset Tmp Offs

 

 

 

 

 

 

 

 

 

 

 

4

Press

Enter

 

 

 

 

 

 

 

 

5

Follow the prompts at the top

“Place probe in sample”

 

and bottom of the screen.

The display will show the

 

 

 

temperature of the reference

 

 

 

solution as measured by the

 

 

 

probe and Analyzer system.

 

 

 

“Press Enter when stable”

 

 

 

6

Once the reading is stable,

“Change to sample value”

 

Press

Enter

Use

to change the value to

 

match the actual temperature of

 

 

 

the reference solution at its

 

 

 

current temperature.

 

 

 

“Enter to save, Exit to cancel”

 

 

 

Limit is ± 5ºC (± 9ºF)

 

 

 

 

7

Press

Enter

This will save the Temperature

 

Offset value. If the calibration is

 

 

 

not successful, an error

 

 

 

message will be displayed.

8If the calibration is suspect, you can reset the Temperature Offset and calibrate again.

January 2009

UDA2182 Universal Dual Analyzer Product Manual

187

Page 199
Image 199
Honeywell UDA2182 manual Place probe in sample, Limit is ± 5ºC ± 9ºF