Input Calibration

Step

Action

Screen

4

Press Enter

“Pressure Sensor Cal”

 

Display shows the barometric

 

Follow the prompts at the top

pressure value in mm Hg.

 

and bottom of the screen.

“Press Enter when stable”

 

 

 

 

 

5

Once the reading is stable,

Change to sample value”

 

press Enter

Use the arrow keys to change

 

the displayed value until the

 

 

displayed pressure in mmHg

 

 

agrees with the known pressure.

 

 

“Enter to save, Exit to cancel”

 

 

 

6

 

“ Cal Complete”

 

 

This screen gives you an option

 

 

to exit or recalibrate.

 

 

Press ENTER to recalibrate.

 

 

Press EXIT to return to Input Cal

 

 

Screen.

 

 

 

7

If the calibration fails, an error

Error Messages:

 

message will be displayed

See Table 12-2.

 

across the bottom stripe of the

 

 

 

screen.

 

 

Make necessary adjustments

 

 

and re-calibrate.

 

 

 

 

Running a Probe Bias Scan

Introduction

The dissolved oxygen probe is an electrochemical cell, which produces an electric current that is directly proportional to the concentration of oxygen dissolved in the sample in which the probe tip is immersed. (When the probe is in air, the current is identical to that produced when the probe is in air-saturated water.) This current is a direct measurement of oxygen level. Usually, the probe is operated at -0.55V with respect to a reference electrode within the probe. (The minus sign is omitted from the screen as well as from the following discussion.) However, in some applications, the performance of the DO probe can be enhanced by using other bias voltages. The purpose of this test is to evaluate whether the probe bias voltage should be adjusted. Possible interference with probe performance may also be inferred from the Probe Bias Test (PBT).

Test initiation

When the test is initiated, the bias voltage is adjusted down from its original value (usually 0.55V) at 25mV/sec until 0V is reached. Then the bias voltage is driven up to 1.0 V at 25mV/sec and finally, it is driven down again until it has returned to the value it had just

172

UDA2182 Universal Dual Analyzer Product Manual

January 2009

Page 184
Image 184
Honeywell UDA2182 manual Running a Probe Bias Scan Introduction, Test initiation, Pressure Sensor Cal