Honeywell UDA2182 Status Parameter Status Definition Display Read Only, Auto Cycling, Configured

Models: UDA2182

1 258
Download 258 pages 46.92 Kb
Page 56
Image 56

Operating the Analyzer

Status

Parameter

Status

Status Definition

Display

 

(Read Only)

 

 

WebPgLngSet

EE/RT/PC

Identifies the web page language set programmed into

 

 

 

the Communications card.

 

 

 

EE web pages support English, French, German,

 

 

 

Italian and Spanish.

 

 

 

RT web pages support English, Russian and Turkish

 

 

 

PC web pages support English, Polish and Czech

 

 

 

 

 

Address

Value

Address states the Modbus RTU slave ID

 

 

 

 

 

Baud Rate

Value

Baud Rate set for RS485

 

 

 

 

 

Word Swap

Yes/No

Word Swap indicates whether the word order set for

 

 

 

Modbus Communications is Little Endian (NO) or Big

 

 

 

Endian (YES)

 

 

 

 

 

MACaddr Hi

Value

MACaddr Hi and MACaddr Low is the MAC address

 

MACaddr Low

 

of the Communication card

 

 

 

 

 

DHCP

Yes/No

DHCP indicates whether Dynamic Host Configuration

 

 

 

Protocol is used.

 

 

 

DHCP is a protocol used by network devices (clients)

 

 

 

to obtain various parameters necessary for the clients

 

 

 

to operate in an Internet Protocol (IP) network.

 

 

 

 

 

IPaddr

Value

IPaddr gives the IP address of the Communication

 

 

 

card.

 

 

 

 

 

SubnetMsk

Value

SubnetMsk indicates the Subnet mask used by the

 

 

 

Communication card.

 

 

 

 

 

Gateway

Value

Gateway Indicates the default Gateway IP address

 

 

 

used by the Communication card.

 

 

 

 

 

DnsSrvr

Value

DnsSrvr displays the DNS (Domain Name Service)

 

 

 

server IP address used by the Communication card

 

 

 

 

 

SMTPsrvr

Value

SMTPsrvr displays the SMTP (Simple Mail Transfer

 

 

 

Protocol) server IP address used by the

 

 

 

Communication card.

 

 

 

 

Auto Cycling

Next Rinse 1

Date and Time

Read Only – The Status Displays menu includes an

(if

Next Rinse 2

 

Auto Cycling selection when any auto cycle is enabled

Next Cal 1

 

and its cycle timer is also enabled. This display

configured)

Next Cal 2

 

provides information on the next occurrences of rinses

 

 

 

and calibrations for any auto cycle with timer

 

 

 

according to the configured cycle count of each

 

 

 

operation.

 

 

 

 

44

UDA2182 Universal Dual Analyzer Product Manual

January 2009

Page 56
Image 56