Configuration

 

 

 

 

 

Sub-menu

Parameter

Selection or

Parameter Definition

selection

 

Range of

 

 

 

 

Setting

 

 

 

TDS Factor

0.010

The TDS Factor is a conversion value applied to

 

(only PV Type

1.000(default)

conductivity to derive total dissolved solids, in units of

 

TDS)

2.000

ppm per μS/cm.

 

 

 

 

 

Temp Type

8550Ω Therm

8550Ω Thermistor

 

 

(default)

1000Ω Resistance Temperature Detector

 

 

1000Ω RTD

 

 

Manual

Manual

 

 

 

 

 

Temp Deg C or F

-10.0 to

If “Manual” is selected at “Temp Type” -Temp Deg F

 

 

140.0ºC

or C will appear depending on what Temperature Unit

 

 

14 to 284ºF

was selected in “Maintenance” setup group,

 

 

parameter “Temp Units”.

 

 

 

 

 

 

 

 

January 2009