Configuration

 

 

 

 

 

Sub-menu

Parameter

Selection or

Parameter Definition

selection

 

Range of

 

 

 

 

Setting

 

 

 

PV Bias

 

PV Bias Constant - is used to compensate the input

 

 

 

for drift of an input value.

 

 

-20.00 to 20.00

If PPM Board is installed.

 

 

PPM

 

 

 

 

-20000 to

If PPB Board is installed.

 

 

20000 PPB

 

 

 

 

default = 0.000

 

 

 

 

 

 

 

Failsafe

 

The output value to which the output will go to protect

 

 

 

against the effects of failure of the equipment.

 

 

0.000 to 20.00

If PPM Board is installed.

 

 

PPM

 

 

 

 

0.000 to 20000

If PPB Board is installed.

 

 

PPB

 

 

 

 

default =

 

 

 

 

20.000

 

 

 

Filter Time

0 to 120.0

A software digital filter is provided for dampening the

 

 

default = 0.0

process. The units are in time constant seconds.

 

 

 

 

 

 

 

 

 

January 2009

UDA2182 Universal Dual Analyzer Product Manual

73