Outputs Calibration

Viewing and resetting 20mA and 4mA Offset

If the calibration is suspect, you can reset the 20mAand 4mA Offset and calibrate again.

In the same screen as “20mA and 4mA Offset”, use the keys to highlight “Reset 20mA Offset” or “Reset 4mA Offset”.

OUTPUT 1

20mA Offset

-147

4mA Offset

3

Reset 20mA Offs

Reset 4mA Offs

Figure 9-1 Resetting Output 1 Offsets (example)

Press ENTER. The 20mA Offset or 4mA Offset will be reset to 0(default).