Operating the Analyzer

The UDA also supports Pharma Europa (PhEur) section 2.2.38, which specifies tests for determining Highly Purified Water which are identical to USP Stages 1, 2 and 3. PhEur adds a less demanding test for determining Purified Water at the end of the sequence.

5.8.3Access to Pharma Display

When Pharma is enabled (see Input Configuration – Section 6.6) press

Display

until you see:

 

 

 

 

The measured temperature-

 

 

 

 

uncompensated conductivity

 

 

 

 

of the solution in µS/cm

Pharma

 

Right Pointer -

 

 

 

 

 

Warning Limit

 

 

 

Measurement

 

 

 

 

relative to test limit

Relay 3 Physical State

Relay 1 Physical State

 

 

 

 

 

 

 

 

White – De-energized

White – De-energized

1

PHARMA 1

 

3

Black - Energized

Black - Energized

 

The measured temperature

High and Pharma Low

 

0.966

25.7 ºC

 

 

 

Right Bar graph – Graphical percent of

Left Bar graph -Measured

 

 

µS/cm

 

of solution in system

solution conductivity scaled by

 

 

 

temperature Units

the parameter values Pharma

 

 

 

 

 

 

 

 

1.30

Stage 1

 

measured conductivity relative to test limit

Left Pointer - Test Limit scaled

 

 

Currently select

Pharma Low

 

74.3%

by parameters Pharma High and

 

 

 

 

 

Pharma Test

 

 

 

 

 

 

Relay 2 Physical State

2

 

 

 

4

The percent of measured

 

 

 

conductivity relative to test limit

White – De-energized

 

 

 

Black - Energized

 

 

 

 

 

 

 

 

Purified water test

 

 

Relay 4 Physical State

 

 

limit in units of

 

 

White – De-energized

 

 

measured value

 

 

Black - Energized

Figure 5-5 Pharma Display screen example

5.8.4Displays

The upper left portion of the screen shows the measured temperature uncompensated conductivity of the solution in uS/cm for Stage 1. For stage 2 and 3 it displays the test conductivity value entered during the measurement in stage 2

The process temperature is shown in the upper right of the screen

The lower right portion of the screen shows the USP stage 1 purified water limit, the stage 2 limit of 2.1 uS/cm and the stage 3 purified water test limit at the measured pH.

When the Pharma n display window is active the various stages can be accessed through the pharma op panel. (Table 5-6)

34

UDA2182 Universal Dual Analyzer Product Manual

January 2009

Page 46
Image 46
Honeywell UDA2182 manual Access to Pharma Display, Pharma Display screen example Displays