Operating the Analyzer

5.10Status Display

Overview

The Status Displays let you see the status of the Alarm Status, PID Alarm Status, Logic Status, Input Status, Output levels, Relay states, Monitor Status, Math Values, Aux Values, Auto Cycling, Variables, Comm Status, System Status, and the Calculated values (Calc Values available only if both units of measurement are identical).

Access to Status Displays

• Press

Display

until you see:

STATUS DISPLAY

Alarm Status

PID Alarm Status

Logic Status

Input Status

Output Levels

Relay States

Monitor Status

Math Values

Aux Values

Variables

Comm Status

Auto Cycling (if configured)

Calc Values (if configured)

 

Use the

keys to highlight the Status Display required.

 

Press Enter

to display the parameters and the status of each.

 

 

Figure 5-8 Status Display screen example

 

 

Table 5-7 Status Display Details

 

 

 

 

 

 

Status

 

Parameter

 

Status

Status Definition

Display

 

 

 

(Read Only)

 

Alarm Status

Alarm 1

 

ON

ON = Latching Alarm in alarm.

 

Alarm 2

 

OFF

Acknowledge alarm by changing status to OFF.

 

Alarm 3

 

 

 

Alarm 4

 

 

If status changes back to ON, alarm condition still

 

 

 

 

 

exists.

 

 

 

 

 

PID Alarm

PID 1 Alm 1

 

ON

ON = PID Alarm Active

Status

PID 1 Alm 2

 

OFF

PID alarms are not latching.

 

PID 2 Alm 1

 

 

 

 

 

 

 

PID 2 Alm 2

 

 

 

 

 

 

 

 

 

January 2009

UDA2182 Universal Dual Analyzer Product Manual

41

Page 53
Image 53
Honeywell UDA2182 manual 10Status Display, Access to Status Displays, Status Display Details, PID Alarm