Unpacking

3 Unpacking, Preparation, and Mounting

3.1 Overview

Introduction

This section contains instructions for unpacking, preparing, and mounting the Analyzer. Instructions for wiring are provided in Section 4 (power wiring) and Section 7 (input wiring). Software configuration is described in Section 6.

The UDA2182 Analyzer can be panel, wall, or pipe mounted.

Each unit has (4) 22.22mm [.87"] dia. holes on the bottom of the unit for lead wires and conduit fittings. The user supplies the conduit fittings.

CAUTION

To avoid damage to the case when connecting to a rigid metallic conduit system, the conduit hub must be connected to the conduit before the hub is connected to the enclosure

ATTENTION

When installing the unit, you must select appropriate watertight fittings to insure watertight integrity.