Diagnostics and Messages

12.4 Auto Cycle Fail Messages

Overview

Auto Cycle Fail is active whenever an auto cycle failure has occurred. The status message “Auto Cycle n Fail” is also displayed during a fail state. Once detected, the current cycle proceeds immediately to the Probe Insert step (if enabled) or to the Resume Delay step. The fail state remains for the duration of the Resume Delay, whereupon the fail state returns to 0 and the fail message is cancelled. A fail state also provides a detail message in the lower half of the Auto Cycle display regarding the specific reason for the error. These messages are listed below:

 

Table 12-3 Auto Cycle Fail Messages

 

 

 

Fail Message

 

Reason

Probe Extract Timeout

 

Probe Transit enabled, Extract Wait Src not None and state not 0

 

 

within Max Transit Mins of start of Probe Extract.

 

 

 

Probe Insert Timeout

 

Probe Transit enabled, Insert Wait Src not None and state not 0

 

 

within Max Transit Mins of start of Probe Insert.

 

 

 

Input Fault

 

Input board, PV or temperature fault has occurred during calibration

 

 

(fault type in status message).

 

 

 

Solution Unstable

 

PH PV or temperature not stable for calibration within elapsed time

 

 

limit Max Cal Mins.

 

 

 

Buffer Span Too Low

 

Difference of PV reading for pH Cal 2 (Slope) and that of last pH Cal

 

 

1 (zero offset) < 1.8 pH.

 

 

 

Offset Underrange

 

PH calibration has calculated and rejected a zero offset < -2 pH.

 

 

 

Offset Overrange

 

PH calibration has calculated and rejected a zero offset > 2 pH.

 

 

 

Slope Underrange

 

PH calibration has calculated and rejected a slope < 80 %.

 

 

 

Slope Overrange

 

PH calibration has calculated and rejected a slope > 105 %.

 

 

 

Readings Unstable

 

DO PV or temperature not stable for calibration within elapsed time

 

 

limit Max Cal Mins.

 

 

 

Probe Current Too Low

 

Probe current is < 5 μA during DO calibration.

 

 

 

194